| Outside Dia of Scope | 8.4mm | ||
| Working length | 1.5 metres | ||
| Insertion Tube
- Distal end, rigid - Bending section |
21.9mm
up 120 degrees, right 100 degrees left 100 degrees |
||
| Bending Radius | 50mm | ||
| Light guide | 2000mm | ||
| Light Sources | 150 Watt halogen and 250 watt halogen | ||
Optical System (Field of view, direction of view, depth of field & illumination system) |
|||
| Forward Viewing lens 30 degrees | 60mm to 250mm | ||
| Forward Viewing lens 60 degrees | 20mm to Infinity | ||
| Forward Viewing lens 100 degrees | 9mm to 200mm | ||
| Side Viewing lens 30 degrees | 45mm to 90mm | ||
| Photo Documentation
- Camera 35mm - 3.5 floppy disc - video |
800ASA J-Peg format Video Super VHS and DV |
||
Electrical Specification for ILK - C 150W Light Source |
|||
| Power Supply AC 240 volt power or
Power input AC 2.4A Power Consumption AC Max 530VA |
DC 12V negative
DC Input 13A(+12V) DC Max 200W |
||
| Frequency | 50/60 Hz | ||
| Power fluctuations AC
DC fluctuations |
+ or - 10%
-0.3V to + 3V |
||
| Operating temperature
- Ambient - Relative humidity - Pressure |
0 - 40 degrees C
15 - 90% (with no condensation) 1013hPa |
||
| Operating Environments of tube
& adaptors
|
|||
| - Insertion tube & optical adaptor
- Other portions except insertion tube |
In air: -10 degrees C to 80 degrees C
In water: -10 degrees C to 80 degrees C In air: -10 degrees C to 50 degrees C |
||
Fluid Resistance |
|||
| - Insertion tube & optical
adaptor
- All portions except insertion tube |
Withstands machine oil, light oil and
5% salt water (10 degrees C to 30 degrees), 1013hPa
Avoid contact (adhering drops) with other fluids than water. |
||
Waterproofness Properties |
|||
| - Insertion tube & optical adaptor
- All portions exept insertion |
Waterproof. Use in fluids possible
Drip proof. May be used under conditions where tube water drops adhere. |
||